The measurement of ocular axial length in normal human eyes based on an improved Twyman‐Green interferometer

Cheng Wang,Shengnan Huangfu,Qi Zhou,Yanrui Li,Huazhong Xiang,Minghui Chen,Gang Zheng,Songlin Zhuang,Jie Zhao,Dawei Zhang
DOI: https://doi.org/10.1002/jbio.202100021
IF: 3.3899
2021-07-12
Journal of Biophotonics
Abstract:<p>In order to meet the demands of myopia prevention, as well as the increasing needs of measurement for refractive and cataract operations in China, a new axial length (AL) measurement system combining an improved Twyman-Green interferometer with digital signal processing has been established. The ALs of 33 eyes of different optical refractive subjects (−8.50 ~ 0.50 D) were measured with the New AL and IOL master. The repeatability of measurements with the New AL was assessed using CoV and ICC. Comparison, correlation, linear regression and agreement of AL between devices were analyzed. There was good repeatability (CoV = 0.0617%, ICC = 0.9999) with the New AL and great agreement has been obtained with both devices. These show that the New AL is capable of providing presice AL values over the normal AL range compared to the IOL master, and indicate that the New AL developed can be used for routine clinical AL measurements.</p><p>This article is protected by copyright. All rights reserved.</p>