Increasing the Sensitivity of pH Glass Electrodes with Constant Potential Coulometry at Zero Current

Robin Nussbaum,Stéphane Jeanneret,Eric Bakker,Stéphane Jeanneret
DOI: https://doi.org/10.1021/acs.analchem.4c00592
IF: 7.4
2024-04-11
Analytical Chemistry
Abstract:It has recently become possible to increase the sensitivity of ion-selective electrodes (ISEs) by imposing a constant cell potential, allowing one to record current spikes with a capacitor placed in series in the circuit. The approach requires a transient current to pass through the measurement cell, which unfortunately may introduce measurement errors and additionally excludes the use of high-impedance indicator electrodes, such as pH glass electrodes. We present here an electronic circuit that...
chemistry, analytical
What problem does this paper attempt to address?