Cross-linkage urease nanoparticles: a high-efficiency signal-generation tag for portable pH meter-based electrochemical immunoassay of lipocalin-2 protein diagnostics

Ai-Li Sun,Qing-An Qi,Li-Juan Zhi
DOI: https://doi.org/10.1007/s00604-020-04466-6
IF: 5.7
2020-08-05
Microchimica Acta
Abstract:An innovative signal-transduction tag based on cross-linked urease nanoparticles (CLENP) was designed for the development of a pH meter-based immunoassay of lipocalin-2 (LCN2). The CLENP was synthesized with a typical desolvation method using ethanol as desolvation agent, followed by functionalization with polyaspartic acid. The carboxylated CLENP were used as the signal-generation tags for the labelling of secondary antibodies via the carbodiimide coupling. Upon target LCN2 introduction, a sandwich-type immune reaction was performed between capture antibody-coated plate and the labeled secondary antibody on the CLENP. The conjugated CLENP in the microplate hydrolyzed urea into ammonia (NH<sub>4</sub><sup>+</sup>) and carbonate (CO<sub>3</sub><sup>2−</sup>), resulting in the pH change of solution, which was determined with a handheld pH meter. The pH variation was proportional to target concentration in the sample. By monitoring the pH variation of the urea solution, the level of LCN2 at a concentration as low as 5.2 pg mL<sup>−1</sup> was evaluated. The pH meter-based electrochemical immunoassay can be utilized for mass production of miniaturized lab-on-a-chip devices with handheld pH meter, thereby opening new opportunities for protein diagnostics and biosecurity.
chemistry, analytical
What problem does this paper attempt to address?