Combining Electrochemiluminescence Detection with Aptamer-Gated Indicator Releasing Mesoporous Nanoparticles Enables ppt Sensitivity for Strip-Based Rapid Tests

Estela Climent,Knut Rurack
DOI: https://doi.org/10.1002/anie.202110744
2021-12-06
Abstract:The combination of electrogenerated chemiluminescence (ECL) and aptamer-gated indicator delivering (gAID) magnetic mesoporous silica nanoparticles embedded into glass fibre paper functionalised with poly(ethyleneglycol) and N-(3-triethoxysilylpropyl)diethanolamine allowed the development of a rapid test that detects penicillin directly in diluted milk down to 50±9 ppt in <5 min. Covalent attachment of the aptamer "cap" to the silica scaffold enabled pore closure through non-covalent electrostatic interactions with surface amino groups, while binding of penicillin led to a folding-up of the aptamer thus releasing the ECL reporter Ru(bpy)32+ previously loaded into the material and letting it be detected after lateral flow by a smartphone camera upon electrochemical excitation with a screen printed electrode inserted into a 3D-printed holder. The approach is simple, generic and presents advantages with respect to sensitivity, measurement uncertainty and robustness compared with conventional fluorescence or electrochemical detection, especially for point-of-need analyses of challenging matrices and analytes at ultra-trace levels.