Matrix overcompensation calibration: a new strategy to correct matrix effects of carbon origin in multielement analysis by inductively coupled plasma mass spectrometry

Guoying Chen,Bunhong Lai
DOI: https://doi.org/10.1016/j.aca.2024.342675
IF: 6.911
2024-05-06
Analytica Chimica Acta
Abstract:Background Elemental analysis by inductively coupled plasma mass spectrometry (ICP-MS) may suffer from matrix effects; those caused by organic matrices cannot be corrected by internal standardization. A new strategy, matrix overcompensation calibration (MOC), was developed to correct such matrix effects. Results Clear fruit juices were diluted 1:50 in 1% HNO 3 (v/v)- 0.5% HCl (v/v)- 5% ethanol (v/v). A standard series was treated likewise to construct an external calibration curve. As, Se, Cd, and Pb in juices were determined by dilute-and-shoot ICP-MS based on this MOC strategy. The results agreed with those obtained by standard addition calibration and microwave-aided digestion; data accuracy was validated by spike-recovery studies. Significance Unlike standard addition calibration, a single external calibration curve established by MOC can be applicable to juices of diversified fruit, geographical, and manufacturer origins enhancing productivity.