Quantitative analysis of Cr in milk powder by LIBS integrated ICCD spectrometer

Ping YANG,Lin HUANG,Ming-yin YAO,Tian-bin CHEN,Hui-qin HU,Cai-hong WANG,Mu-hua LIU
DOI: https://doi.org/10.13836/j.jjau.2015141
2015-01-01
Abstract:The experiment is to improve the accuracy of laser-induced breakdown spectroscopy (LIBS) to detect heavy metal Cr element in milk powder,with integrated spectrometer ICCD plasma signals in milk sam-ples were collected.To reduce detection errors,first the spectral line profiles were pretreated by means of the Lorentz fit,extracting the net spectral intensity.Under the same test conditions,using CrI 425.43 nm as analyti-cal lines and CaI 422.67 nm as the internal standard line,using the external standard method and the internal standard method respectively the quantitative analysis of elements Cr in milk powder was conducted,and the linear coefficients of the two kinds of calibration curves were 0.946 52 and 0.991.The experiment result shows that,the selecting matrix elements Ca in milk powder as an internal standard substance can effectively improve the calibration curve fitting effect and reduce errors,which provides a reference for using the standard modeling method in portable LIBS technique to food detection in future.
What problem does this paper attempt to address?