Clinical Evaluation of a New Spectral-Domain Optical Coherence Tomography-Based Biometer

Jorge L. Alió,Marina José-Martínez,Antonio Martínez-Abad,Alejandra E. Rodríguez,Francesco Versaci,Jesper Hjortdal,Joaquim Neto Murta,Ana B. Plaza-Puche,Mario Cantó-Cerdán,David P. Piñero
DOI: https://doi.org/10.3390/diagnostics14050560
IF: 3.6
2024-03-07
Diagnostics
Abstract:The VEMoS-AXL system is a new optical biometer based on spectral domain optical coherence tomography (SD-OCT) that has been tested in terms of intrasession repeatability and compared with a swept-source optical coherence tomography biometer (SS-OCT), which is recognized as the gold standard for the performance of an agreement analysis. A biometric analysis was performed three consecutive times in 120 healthy eyes of 120 patients aged between 18 and 40 years with the SD-OCT system, and afterwards, a single measurement was obtained with the SS-OCT system. Within-subject standard deviations were 0.004 mm, 4.394 μm, and 0.017 mm for axial length (AL), central corneal thickness (CCT), and anterior chamber depth (ACD) measures obtained with the SD-OCT biometer, respectively. The agreement between devices was good for AL (limits of agreement, LoA: −0.04 to 0.03 mm) and CCT (LoA: −4.36 to 14.38 μm), whereas differences between devices were clinically relevant for ACD (LoA: 0.03 to 0.21 mm). In conclusion, the VEMoS-AXL system provides consistent measures of anatomical parameters, being most of them interchangeable with those provided by the SS-OCT-based gold standard.
medicine, general & internal
What problem does this paper attempt to address?