Evaluation of a new dynamic real-time visualization 25 kHz swept-source optical coherence tomography based biometer

Jinjin Yu,Xuanqiao Lin,Xiaomin Huang,Zhenyu Xu,Rui Ning,Kexin Li,Giacomo Savini,Domenico Schiano-Lomoriello,Xingtao Zhou,Jinhai Huang,Xingtao Zhou and Jinhai Huang
DOI: https://doi.org/10.1186/s40662-024-00377-2
2024-03-05
Eye and Vision
Abstract:To evaluate the intraobserver repeatability and interobserver reproducibility of a newly developed dynamic real-time visualization 25 kHz swept-source optical coherence tomography (SS-OCT) based biometer (ZW-30, TowardPi Medical Technology Ltd, China) and compare its agreement with another SS-OCT based biometer (IOLMaster 700, Carl Zeiss Meditec AG, Jena, Germany).
ophthalmology
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to evaluate the intra - observer repeatability and inter - observer reproducibility of a new type of dynamic real - time visual 25 kHz swept - source optical coherence tomography (SS - OCT) biometer (ZW - 30), and compare its measurement consistency with another SS - OCT - based biometer (IOLMaster 700). Specifically, the study aims to: 1. **Evaluate the performance of the new biometer (ZW - 30)**: - **Intra - observer repeatability**: Evaluate the consistency of measurement results of the same operator at different time points by measuring the same parameter multiple times. - **Inter - observer reproducibility**: Evaluate the consistency of measurement results among different operators by the measurement results of different operators for the same parameter. 2. **Compare the measurement consistency of the two biometers**: - Use Bland - Altman plots to evaluate the consistency of the two devices when measuring parameters such as axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), mean corneal curvature (Km), astigmatism (AST), J0 vector, J45 vector and corneal diameter (CD). ### Main research methods - **Research subjects**: 82 patients with healthy right eyes. - **Measurement devices**: New biometer (ZW - 30) and IOLMaster 700. - **Measurement parameters**: Including AL, CCT, ACD, LT, Km, AST, J0, J45 and CD. - **Statistical analysis**: Use SPSS and Excel software for data analysis, including Kolmogorov - Smirnov test, paired t - test, Bland - Altman plots, etc. ### Research results - **Intra - observer repeatability and inter - observer reproducibility**: - The new biometer (ZW - 30) shows excellent intra - observer repeatability and inter - observer reproducibility in all parameters, with low Sw, TRT and CoV values, and ICC values close to 1.000. - Only the ICC value of J45 is relatively low (0.887 - 0.899). - **Measurement consistency between the two devices**: - Although there are statistically significant differences (P < 0.05) in AL, AQD, ACD, Km, J45 and CD values, the 95% limit of agreement (LoA) is relatively narrow. - The measurement results of CCT, LT, AST and J0 values between the two devices are similar, and the maximum absolute 95% LoA are 8.08 μm, 0.08 mm, 0.30 D and 0.16 D respectively. ### Conclusions - The new dynamic real - time visual biometer (ZW - 30) shows excellent intra - observer repeatability and inter - observer reproducibility when measuring ocular parameters. - ZW - 30 and IOLMaster 700 have a high degree of consistency when measuring ocular parameters and can be used interchangeably in clinical practice. ### Formulas - **J0 vector**: \[ J0 = -\frac{(Ks - Kf)}{2} \times \cos(2\alpha) \] - **J45 vector**: \[ J45 = -\frac{(Ks - Kf)}{2} \times \sin(2\alpha) \] - where \( Ks \) and \( Kf \) represent the steepest and flattest corneal curvatures respectively, and \( \alpha \) represents the axis of the cylinder lens. Through these evaluations, the study provides strong evidence for the reliability and accuracy of the new biometer (ZW - 30) in clinical applications.