Single-sided magnetic resonance-based sensor for point-of-care evaluation of muscle

Sydney E. Sherman,Alexa S. Zammit,Won-Seok Heo,Matthew S. Rosen,Michael J. Cima
DOI: https://doi.org/10.1038/s41467-023-44561-9
IF: 16.6
2024-01-10
Nature Communications
Abstract:Abstract Magnetic resonance imaging is a widespread clinical tool for the detection of soft tissue morphology and pathology. However, the clinical deployment of magnetic resonance imaging scanners is ultimately limited by size, cost, and space constraints. Here, we discuss the design and performance of a low-field single-sided magnetic resonance sensor intended for point-of-care evaluation of skeletal muscle in vivo. The 11 kg sensor has a penetration depth of >8 mm, which allows for an accurate analysis of muscle tissue and can avoid signal from more proximal layers, including subcutaneous adipose tissue. Low operational power and shielding requirements are achieved through the design of a permanent magnet array and surface transceiver coil. The sensor can acquire high signal-to-noise measurements in minutes, making it practical as a point-of-care tool for many quantitative diagnostic measurements, including T2 relaxometry. In this work, we present the in vitro and human in vivo performance of the device for muscle tissue evaluation.
multidisciplinary sciences
What problem does this paper attempt to address?
The problem that this paper attempts to solve is: how to design and implement a portable, low - field - strength single - sided magnetic resonance (SSMR) sensor for rapid and accurate assessment of skeletal muscles in the point - of - care (POC) environment. ### Problem Background Although traditional magnetic resonance imaging (MRI) performs well in soft - tissue pathological detection, due to its large size, high cost and space limitations, it is difficult to be widely applied in the point - of - care environment. In addition, traditional MRI requires a high magnetic field strength (such as 1.5 - 3 Tesla), which brings safety risks (for example, the ejection risk of ferromagnetic objects), and requires strict shielding and power supply conditions, further limiting its application range. ### Paper Goals To solve the above problems, this research has designed a lightweight (11 kg), low - field - strength (0.2 T) single - sided magnetic resonance sensor specifically for the assessment of skeletal muscles in the point - of - care environment. This sensor has the following characteristics: - **Penetration Depth**: > 8 mm, which can avoid the subcutaneous fat layer and directly detect muscle tissue. - **Low Power Consumption and Shielding Requirements**: By designing a permanent magnet array and a surface transceiver coil, low power consumption and minimized shielding requirements are achieved. - **Rapid Measurement**: It can obtain high signal - to - noise ratio (SNR) measurement results within a few minutes, making it suitable for a variety of quantitative diagnostic measurements, such as T2 relaxation measurement. ### Solutions To achieve these goals, the researchers took the following measures: 1. **Magnet Array Design**: Finite element analysis was carried out using COMSOL Multiphysics software to optimize the design of the magnet array to ensure the formation of a uniform magnetic field area at a distance of 8 mm from the sensor surface. 2. **Signal Acquisition and Processing**: The Car - Purcell - Meiboom - Gill (CPMG) pulse sequence was used for signal acquisition to maximize the signal - to - noise ratio and avoid the influence of B0 magnetic field inhomogeneity. 3. **Thermal Management and Grounding Design**: The RF coil was wrapped with aluminum nitride (AlN) sheets for effective heat dissipation to prevent the coil from overheating; at the same time, a reliable grounding system was designed to reduce the influence of external electromagnetic interference (EMI). ### Experimental Verification The researchers verified the effectiveness of the sensor through in - vitro experiments (using a multi - layer tissue model) and in - vivo experiments (testing on healthy volunteers). The results showed that the sensor can obtain high - quality muscle signals in a relatively short time and can distinguish between muscle and subcutaneous fat tissue. ### Potential Applications This new SSMR sensor is expected to provide rapid and accurate muscle assessment in the point - of - care environment, especially in emergency, intensive care and other scenarios, helping doctors make diagnosis and treatment decisions more quickly. ### Conclusion This research has successfully designed and verified a single - sided magnetic resonance sensor suitable for the point - of - care environment, solving the limitations of traditional MRI in terms of portability and application range, and providing a new tool for clinical diagnosis.