CdTe XG‐Cam: A new high‐resolution x‐ray and gamma‐ray camera for studies of the pharmacokinetics of radiopharmaceuticals in small animals

Miho Katsuragawa,Atsushi Yagishita,Shin'ichiro Takeda,Takahiro Minami,Kazunobu Ohnuki,Hirofumi Fujii,Tadayuki Takahashi
DOI: https://doi.org/10.1002/mp.17124
IF: 4.506
2024-05-21
Medical Physics
Abstract:Background The recent emergence of targeted radionuclide therapy has increased the demand for imagers capable of visualizing pharmacokinetics in developing radiopharmaceuticals in the preclinical phase. Some radionuclides emit hard x‐rays and gamma‐rays below 100 keV, in which energy range the performance of conventional NaI scintillators is poor. Multipinhole collimators are also used for small animal imaging with a good spatial resolution but have a limited field of view (FOV). Purpose In this study, a new imager with high sensitivity over a wide FOV in the low‐energy band (
radiology, nuclear medicine & medical imaging