An Amplitude-Programmable Energy-Recycling High-Voltage Resonant Pulser for Battery-Powered Ultrasound Devices

Imad Bellouki,Nuriel N. M. Rozsa,Zu-Yao Chang,Zhao Chen,Mingliang Tan,Michiel A. P. Pertijs
DOI: https://doi.org/10.1109/jssc.2024.3494536
IF: 5.4
2024-01-01
IEEE Journal of Solid-State Circuits
Abstract:This article presents an application-specific integrated circuit (ASIC) for battery-powered ultrasound (US) devices. The ASIC implements a novel energy-efficient high-voltage (HV) pulser that generates HV transmit (TX) pulses directly from a low-voltage (LV) battery supply. By means of a single off-chip inductor, energy is supplied to a US transducer in a resonant fashion, directly generating half-period sinusoidal HV pulses on the transducer, while consuming substantially less energy than a conventional class-D pulser. By recycling residual reactive energy from the transducer back to the input, the energy consumption is further reduced by more than 50%. The autocalibration techniques are leveraged to deal with tolerances of the inductor, transducer, and battery supply and thus maximize the energy efficiency. A prototype chip was fabricated in TSMC 0.18- $\mu$ m HV BCD technology and used to drive external 120-pF capacitive micromachined US transducers (CMUTs) with a center frequency of approximately 2.5 MHz. Electrical measurements show that the prototype can generate pulses with a peak amplitude between 10 and 30 V accurate to within $\pm$ 1 V. Acoustic measurements demonstrate successful ultrasonic pulse transmission and pulse-echo measurements. The prototype reaches a peak efficiency of 0.23 fCV ${2}$ , which is the highest reported to date for HV pulsers targeting US imaging.
What problem does this paper attempt to address?