Development of Spectral Radiation Imaging Diagnostic Systems for HFRC Device

Cheng Han,Zhifeng Cheng,Weijie Lin,Ye Tian,Yuchen Zhang,Bo Rao,Xinyi Jin
DOI: https://doi.org/10.1088/1742-6596/2849/1/012038
2024-01-01
Journal of Physics Conference Series
Abstract:Abstract Spectral radiation imaging diagnostic systems utilizing PDA (avalanche PhotoDiode Array) have been developed and deployed on the HFRC device (HUST Field Reversed Configuration Device). These systems are designed to measure Hα radiation imaging, plasma velocity, orientation, and shape. Their innovative features include higher spatial resolution, integrated system design, flexible measurement, and superior electronic performance. This paper discusses the selection of basic electronic circuits, circuit analysis, parameter design, and noise analysis in detail. The accuracy of the design is validated through simulation and practical testing. The final system parameters achieved a bandwidth of 610 kHz, low noise of 2.5 mV, approximately 3% system overshoot, and a rising time of 468.5 ns. Additionally, preliminary measurement results are provided.
What problem does this paper attempt to address?