Design and realization of an long-wave infrared hyperspectral imaging system

Liyin Yuan,Ying Lin,Zhiping He,Weiming Xu,Yingqing Zhang,Rong Shu,Jianyu Wang
DOI: https://doi.org/10.3969/j.issn.1007-2276.2011.02.001
2011-01-01
Abstract:In view of the strong background radiation and the low signal noise rate of the long-wave hyperspectral imaging spectrometer, an infrared spectral imaging system which could restrain the background radiation was designed. The background radiation was analyzed by the TacePro software. The analysis include composition of the background radiation (within the whole spectral range), relative background radiation of each channel, background radiation as different mechanism inwall surfaces absorbance, and background radiation at different temperature of machine inwall. The background radiation was mainly suppressed by cooling opto-mechanic system and polishing inwall surface. A ground-based experimental device was established, whose spectral range was from 7.7 to 9.3 μm, spectral resolution was 54 nm, spatial resolution was 0.75 mrad and scanning way was push broom. Measurement of the whole instrument presents that spectral resolution (SRF) of the system reaches the designed value and NETD is less than 300 mK as the inwall surfaces of opto-mechanic is at 150 K.