Teaching the concepts of Donnan potential and liquid-junction potential by evaluating homemade pH and ion-sensitive probes constructed using commercial ion-exchange membranes and reference electrodes

Shane Ardo,William Gaieck,Anni Zhang,Ashley Sabatose,Madison Ngo,Eric Schwartz,William White,Simon Luo,Leanna Schulte,Yixian Wang
DOI: https://doi.org/10.26434/chemrxiv-2024-hxgnv
2024-10-21
Abstract:Transport of ions across semipermeable membranes, like ion-exchange membranes, and salt bridges, like porous frits, each results in electric potential differences that are sensed in most electrochemical experiments. While often neglected, these electric potential differences can significantly impact the interpretation of experimental results. To assist in comprehension of these concepts, a one-to-two-hours-long hands-on laboratory activity, intended for students in electroanalytical chemistry courses, was designed, implemented, and evaluated. This activity requires that students construct a simple two-compartment electrochemical cell from inexpensive and readily available disposable plastic cuvettes, commercially available polymeric ion-exchange membranes, two nominally identical reference electrodes placed in liquid electrolytes that wet each side of the membrane, and a voltmeter or a potentiostat to measure the open-circuit potential between the reference electrodes, i.e. the cell potential. Using commercially available reference electrodes, each consisting of a wire (e.g. Ag/AgCl) immersed in a fritted tube containing an aqueous electrolyte, measured open-circuit potentials report on the magnitude of both Donnan and liquid-junction electric potential differences. These electric potentials are sensitive to the composition of ions in each electrolyte, through interactions with the ion-exchange membrane and porous frits at the tip of each reference electrode. When a cation-exchange membrane separates aqueous solutions consisting of different concentrations of hydrochloric acid (HCl) or potassium hydroxide (KOH), the cell responds like a pH or pK probe, respectively, because the membrane interfaces are more sensitive to the activity of H+ and K+ over Cl– and OH–, respectively. Analogously, when an anion-exchange membrane is used, the cell responds like a pCl or pOH probe, respectively, because the membrane interfaces are more sensitive to the respective activity of Cl– and OH– over H+ and K+. When the solutions contain large concentrations of multiple types of ions, measured open-circuit potentials typically deviate from these Donnan-dominated effects and are instead dominated by a liquid-junction-like effect where electric potentials within and across ion-exchange membranes form due to differences in the permeabilities of the ions. When reference electrodes consist of wires immersed directly into the aqueous electrolytes that wet the membranes, and not in fritted tubes, measured open-circuit potentials also include contributions from the chemical potential of redox-active species that are present in the aqueous electrolytes. While this hands-on laboratory activity is suitable for most upper-level undergraduate students in physical science and engineering disciplines, the concepts of Donnan electric potential, liquid-junction electric potential, and species chemical potential are advanced enough to educate graduate students and senior researchers alike. Our activity also helps instructors teach a topic that is pertinent to the electrochemistry of many aqueous chemical systems using simple components and a straightforward setup.
Chemistry
What problem does this paper attempt to address?
The problem this paper attempts to address is to help students understand the concepts of Donnan potential and liquid junction potential by constructing and evaluating homemade pH probes and ion-sensitive probes. Specifically, the author designed a 1 to 2-hour hands-on experimental activity aimed at enabling students to understand and master these complex electrochemical concepts through practical operation in an electroanalytical chemistry course. ### Main Issues: 1. **Understanding Donnan Potential**: Donnan potential is the potential difference formed at the interface between an ion-exchange membrane and a solution, caused by certain ions being unable to pass through the membrane. This potential difference affects ion transport and electrochemical equilibrium in the solution. 2. **Understanding Liquid Junction Potential**: Liquid junction potential is the potential difference caused by the difference in ion permeability between different electrolyte solutions. This potential difference also affects the interpretation of electrochemical measurement results. 3. **Experimental Teaching**: Through simple experimental setups (such as homemade two-chamber electrochemical cells), students can construct and test these probes by themselves, thereby better understanding these concepts. ### Experimental Methods: - **Experimental Setup**: Construct a simple two-chamber electrochemical cell using inexpensive and readily available materials (such as disposable plastic cuvettes, commercial ion-exchange membranes, reference electrodes, and a voltmeter or potentiometer). - **Experimental Steps**: Add acid or base solutions of different concentrations to the two chambers, measure the open-circuit potential between the reference electrodes to evaluate the effects of Donnan potential and liquid junction potential. - **Specific Experiments**: - **pH Probe**: Observe the change in open-circuit potential by changing the acid concentration in one chamber to verify the formation of Donnan potential. - **pK Probe**: Use potassium salts instead of acids to observe the effect of potassium ions on the open-circuit potential. - **Leakage Probe**: Use different types of electrolyte solutions to demonstrate the effect of multiple ions on potential measurement. - **Other Probes**: Further explore the selectivity and potential response of other ions by replacing different ion-exchange membranes. ### Safety Considerations: - When using diluted acid and base solutions, appropriate personal protective equipment should be worn to prevent skin contact, eye contact, and inhalation. - After the experiment, the remaining solutions should be disposed of according to local regulations. - When using hydrogen reference electrodes, the experiment should be conducted in a well-ventilated environment to prevent hydrogen leakage and potential explosion. Through this experimental activity, students can not only intuitively understand the concepts of Donnan potential and liquid junction potential but also improve their practical skills in electrochemical experiments.