Ultrasensitive, Self-Calibrated Cavity Ring-Down Spectrometer for Quantitative Trace Gas Analysis.

Bing Chen,Yu‐Xi Sun,Ze‐Hua Zhou,Jian Chen,Anwen Liu,S.-M. Hu
DOI: https://doi.org/10.1364/ao.53.007716
IF: 1.9
2014-01-01
Applied Optics
Abstract:A cavity ring-down spectrometer is built for trace gas detection using telecom distributed feedback (DFB) diode lasers. The longitudinal modes of the ring-down cavity are used as frequency markers without active-locking either the laser or the high-finesse cavity. A control scheme is applied to scan the DFB laser frequency, matching the cavity modes one by one in sequence and resulting in a correct index at each recorded spectral data point, which allows us to calibrate the spectrum with a relative frequency precision of 0.06 MHz. Besides the frequency precision of the spectrometer, a sensitivity (noise-equivalent absorption) of 4×10−11 cm−1 Hz−1/2 has also been demonstrated. A minimum detectable absorption coefficient of 5×10−12 cm−1 has been obtained by averaging about 100 spectra recorded in 2 h. The quantitative accuracy is tested by measuring the CO2 concentrations in N2 samples prepared by the gravimetric method, and the relative deviation is less than 0.3%. The trace detection capability is demonstrated by detecting CO2 of ppbv-level concentrations in a high-purity nitrogen gas sample. Simple structure, high sensitivity, and good accuracy make the instrument very suitable for quantitative trace gas analysis.