Ultrasensitive, self-calibrated cavity ring-down spectrometer for quantitative trace gas analysis.

Bing Chen,Yu R Sun,Ze-Yi Zhou,Jian Chen,An-Wen Liu,Shui-Ming Hu
DOI: https://doi.org/10.1364/AO.53.007716
IF: 1.9
2014-01-01
Applied Optics
Abstract:A cavity ring-down spectrometer is built for trace gas detection using telecom distributed feedback (DFB) diode lasers. The longitudinal modes of the ring-down cavity are used as frequency markers without active-locking either the laser or the high-finesse cavity. A control scheme is applied to scan the DFB laser frequency, matching the cavity modes one by one in sequence and resulting in a correct index at each recorded spectral data point, which allows us to calibrate the spectrum with a relative frequency precision of 0.06 MHz. Besides the frequency precision of the spectrometer, a sensitivity (noise-equivalent absorption) of 4 x 10(-11) cm(-1) Hz(-1/2) has also been demonstrated. A minimum detectable absorption coefficient of 5 x 10(-12) cm(-1) has been obtained by averaging about 100 spectra recorded in 2 h. The quantitative accuracy is tested by measuring the CO2 concentrations in N-2 samples prepared by the gravimetric method, and the relative deviation is less than 0.3%. The trace detection capability is demonstrated by detecting CO2 of ppbv-level concentrations in a high-purity nitrogen gas sample. Simple structure, high sensitivity, and good accuracy make the instrument very suitable for quantitative trace gas analysis. (C) 2014 Optical Society of America