Abstract:This article reports on the development of an atmospheric pressure filtering electrospray ionization (FESI) source - composed of a syringe, a filter, and the ESI source - for the determination of furazolidone, amoxicillin, nitrofurantoin, and nitrofural, common antibiotics. Spiked food samples were also analyzed; the analytes were placed in the filter head, and the flow rate of the extraction solution from the syringe pump was adjusted. Using a mixture of oleic acid and antibiotics, hydrophilic and hydrophobic filters were compared, demonstrating the advantage of FESI in reducing the background and noise for the mixture. The device simplifies not only the experimental procedure but also the sample pretreatment process. Additionally, the limits of detection of the antibiotics were determined to be 1 mg L-1 for furazolidone, amoxicillin, and nitrofurantoin and 2 mg L-1 for nitrofural. The linear regression coefficients (R2) of the antibiotic standard solutions were all >= 0.99. Furthermore, the spiked recovery rates of pork, chicken, and egg samples ranged from 73.3% to 107.5%, with relative standard deviations (RSDs) from 1.1% to 15.6%. This study uncovers the potential capabilities of ion trap mass spectrometry (ITMS) for the on-site detection of antibiotics and offers a new method for ensuring food safety.
What problem does this paper attempt to address?