Comparison of Abbott and Da-an Real-Time PCR for Quantitating Serum HBV DNA

Ning,Qiu,Rui,Qing Li,Jianguo,Anchi Yu,Charles H.-P. Wen,Kan Yang,Hsi-Hsien Wei,Bo Zhang,Yong,An,Tong,Xue-en,Ming Liu,Hui,Zhuang zhuang
DOI: https://doi.org/10.3748/wjg.v20.i33.11762
2014-01-01
Abstract:AIM: To compare the performance of the Da-an realtime hepatitis B virus(HBV) DNA assay and Abbott RealTime HBV assay. METHODS: HBV DNA standards as well as a total of 180 clinical serum samples from patients with chronic hepatitis B were measured using the Abbott and Daan real-time polymerase chain reaction(PCR) assays. Correlation and Bland-Altman plot analysis was used to compare the performance of the Abbott and Daan assays. The HBV DNA levels were logarithmically transformed for analysis. All statistical analyses were performed using SPSS for Windows version 18.0. Thecorrelation between the two assays was analyzed by Pearson’s correlation and linear regression. The BlandAltman plots were used for the analysis of agreement between the two assays. A P value of < 0.05 was considered statistically significant.RESULTS: The HBV DNA values measured by the Abbott or Da-an assay were significantly correlated with the expected values of HBV DNA standards(r = 0.999, for Abbott; r = 0.987, for Da-an, P < 0.001). A BlandAltman plot showed good agreement between these two assays in detecting HBV DNA standards. Among the 180 clinical serum samples, 126 were quantifiable by both assays. Fifty-two samples were detectable by the Abbott assay but below the detection limit of the Da-an assay. Moreover, HBV DNA levels measured by the Abbott assay were significantly higher than those of the Da-an assay(6.23 ± 1.76 log IU/mL vs 5.46 ± 1.55 log IU/mL, P < 0.001). A positive correlation was observed between HBV DNA concentrations determined by the two assays in 126 paired samples(r = 0.648, P < 0.001). One hundred and fifteen of 126(91.3%) specimens tested with both assays were within mean difference ± 1.96 SD of HBV DNA levels.CONCLUSION: The Da-an assay presented lower sensitivity and a narrower linear range as compared to the Abbott assay, suggesting the need to be improved.
What problem does this paper attempt to address?