Analysis of Intracranial Aneurysm Embolization with MDS and with GDC

Meng Lv
2001-01-01
Abstract:Objective To evaluate the MDS(Mechanical Detachable Spiral)/GDC (Guglielmi Detachable Coil) embolization effects of intracranial saccular aneurysms. Methods Among 120 cases of intracranial aneurysm (125) during March 1995 to July 1999, 65 cases (66 aneurysms) were embolized with MDS, 48 (51) with GDC and 8 with a combination of MDS and GDC. Clinical material including sex, age, SAH and Hunt Hess Scale, diameter and neck width of aneurysm, the number and length of coils used per aneurysm, occlusive ratio and complications were compared between MDS and GDC groups. Results MDS and GDC groups were comparable (P0 10) in terms of age, sex, diameter of aneurysm(8 46±3 42mm verse 7 38±3 45mm), neck width (3 49±1 50mm NS 3 26±1 52mm). Among those occluded ≥80%, MDS verse GDC: the average occlusive ratios, 95 00%±6 32∶94 19%±7 63%;permanent complications (including death), 7 8% NS 4 2%;Average coils and length per aneurysm, 4 65±3 04 NS 4 24±2 65 and 460 2±398 5mm NS 422 9±387 1mm. Conclusions Although no significant differeace was found when compared MDS with GDC, GDC is more preferable regarding its good compliance, alternative types ready for use, tractability, especially for early ruptured and wide neck aneurysms. Comparably, MDS is not promising as to its difficulty of control, possible propensity of complication, and potentiality of dissolution in spite of its good thrombogenesis and inexpensiveness.
What problem does this paper attempt to address?