Single Venous Pathway for Transcatheter Closure of Children Patent Ductus Arteriosus with Native Made Occluder

Wei Zhang
2006-01-01
Abstract:Objective To evaluate the efficiency and safety in the application of native made occluder for single-verous transcatheter closure of children patent ductus atrenious (PDA). Methods 12 Patients aged 1-12(3.8±4.6) years with body weight 9.5-35.0(12.6±8.7)kg were established RA→RV→PA→PDA path through femoral vein. The PDA diameter at narrowest segment was 3~12(5.35±3.26)mm, according to the lateral descending aortographies. The suitable native made occluder was selected. Echocardiographies and heart auscultation were performed to evaluate the immediate results the procedure. Echocardiographies were performed 24 hours, 1 month, 3 month, 6 month after the procedure to find whether there was residual shunt and recanalization. Results 12 patients had successful device placement at the first time. The success rate was 100%. The pulmonic systole pressure revived to normal (20.30±3.10)mm Hg after the procedure .Continuous murmur disappeared in all patients. The fluoroscopy time was 7.50-12.30(8.96±2.30)min. There were no complications. No residual shunt and PDA recanalization after the complete closure were found during 1, 3 and 6 months follow-up. Conclusion The single-verous transcatheter closuer of PDA using native made occluder device simplify operation procedure, shorten the fluoroscopy time. Similar results was got by single-verous transcatheter closure with traditional operational precedure.
What problem does this paper attempt to address?