Transcatheter Closure of Patent Ductus Arteriosus Using Amplatzer Occluder Device

Zhou Jun,Huili Li
2003-01-01
Abstract:Objective To evaluate the efficiency of transcatheter closure of patent ductus arteriosus(PDA) using Amplatzer occluder device. Methods Twenty two patients with PDA were treated by transcatheter closure using Amplatzer occluder device.All patients were diagnosed definitely by transthoracic echocardiography(TTE) before the procedure. TTE was performed at 3 day and 3 mnths after the procedure, respectively. Results No continuous murmur was heard in all patients instantly after the procedure. Aortography showed that the lesions were closed completely in 18 cases (81.8%), only a little residual shunt remained in four cases 15 minutes after the procedure. TTE revealed that only one patient had a little residual shunt at 3 months after procedure,and disappeared at one year. No complication was found in all patients except one who had pericardial effusion. In a follow up of 3 months, no device displacement, recanalization of ductus arteriosus and pulmonary stenosis appeared. Conclusion Transcatheter closure of PDA using Amplatzer occluder device has high rate of success,wide indication, and little trauma. Therefore,it is an ideal procedure in treatment of PDA at present.