A 128-electrodes Neural Probe with 30*55 µm<sup>2</sup> Channel Area Low-power CCO-based ADC

Weixiong Qiu,Shihui Sun,Yufei Ai,Wengao Lu,Yacong Zhang,Zhongjian Chen
DOI: https://doi.org/10.1109/ASICON58565.2023.10396234
2023-01-01
Abstract:This paper presents a CMOS neural probe that integrates a CCO-based ADC under each electrode with pipeline readout process. CCO-based ADC using a highly digital approach to realizing analog functionality can bring benefit to smaller the area and be more compatible with CMOS process scaling. Preprocessing the neural signal in channel avoid noise-sensitive analog signal long-distance transmission. Each probe integrates 128 electrodes on the shank with a width of 80 µm and a length of 4000 µm. Each electrode occupies an area of 55 * 30 µm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> and consumes a power of 8.58 µW with a total input-referred noise of 3.80 µV <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">rms</inf> .
What problem does this paper attempt to address?