Digital Controlled Gradient Amplifier for Magnetic Resonance Imaging

Xiaohua Jiang
2010-01-01
Abstract:A switch-mode gradient amplifier based on dual parallel H-bridge topology was presented.Dynamic dead-time control method was proposed to reduce the reverse recovery current of power MOSFET.The output filter was designed and optimized for both common and differential mode with good damping and small losses.Further a high performance current tracking controller,combing with state feedback,feed forward and nonlinear PI regulations,was proposed based on the decoupling of common and differential mode of the gradient amplifier model.To ease the requirements of sampling system,a simple PI controller was applied instead of the tracking controller during the steady period.Finally,a dual-DSP control system was designed to implement the control algorithm,and a 200 A/150 V gradient amplifier was built in the laboratory.Experimental results show that only 1% overshoot occurs when tracking a trapezium current with 200 A amplitude and 200 μs rising time;the low frequency noise is in the order of mA when outputing DC current.
What problem does this paper attempt to address?