Research on Two-Level Gradient Amplifier for Magnetic Resonance Imaging System

JIANG Dan-dan,LI Si-qi,JIANG Xiao-hua
DOI: https://doi.org/10.3969/j.issn.1000-100x.2006.06.038
IF: 5.967
2006-01-01
IEEE Transactions on Power Electronics
Abstract:The gradient amplifier is an important equipment of Magnetic Resonance Imaging System(MRI).In this pa-per,a two-level topology gradient amplifier is designed for our MRI gradient system,some simulation and experiment re-search on the topology are done.The result of simulation and experiment show the two-level topology is suitable for MRI gradient system,it can achieve fast transit and high precision of the trapezium shaped current pulses,and a novel PWM control strategy for reducing the output current ripple is suggested.
What problem does this paper attempt to address?