Highly Integrated Gradient Pulse Generator for Magnetic Resonance Imaging System

Weinan Tang,Weimin Wang
DOI: https://doi.org/10.1002/cmr.b.20190
2011-01-01
Abstract:A gradient pulse generator for magnetic resonance imaging is presented. It has the ability to yield arbitrary gradient pulses on the fly and provide reconfigurable calculations including scaling, axis rotation, pre-emphasis, and gradient offsetting in real time along X, Y, and Z gradient channels. The versatility has been obtained by incorporating all required digital functions into a single field programmable gate array chip. The development of the hardware is eased by employing a high-level tool-the System Generator-using the MathWorks model-based Simulink environment. In addition, an easy-to-use method is developed for pre-emphasis calibration. As expected, the described device is highly integrated, low cost, simple to use and interface, and has been demonstrated successfully in conjunction with a home-built magnetic resonance system to perform imaging experiments. (C) 2011 Wiley Periodicals, Inc. Concepts Magn Reson Part B ( Magn Reson Engineering) 39B: 59-63, 2011
What problem does this paper attempt to address?