Dual Transverse Fiber Tip Fabry-Perot Cavities for Simultaneous Measurements of Pressure and Temperature

Shuo Zhang,Xun Wu,Haijie He,Zhangwei Yu,Wei Chen,Nan Li,Erik Forsberg,Shengnan Wu,Sailing He
DOI: https://doi.org/10.1016/j.yofte.2024.103666
IF: 2.8
2024-01-01
Optical Fiber Technology
Abstract:A fiber tip sensor based on dual transverse Fabry-Perot (FP) cavities for simultaneous temperature and gas pressure measurements is proposed, fabricated and tested. The proposed fiber tip sensor is formed by one-step polishing the end face of a microstructure fiber, which contains a hollow hole in the center and an eccentric core, to 45 -degrees. The hybrid fiber tip sensor consists of a quartz cavity and a gas cavity effectively solving the problem of cross sensitivity in dual -parameter sensors. By tracking the shift of the interference spectra of the two FP cavities, simultaneous measurements of temperature and pressure is achieved. With advantages of a compact structure, high spatial resolution, simple fabrication and ability of multi -parameter detection, the sensor is suitable for use in a high temperature and high-pressure simultaneous sensing applications. Furthermore, the fabrication approach ensures consistent size of the FP cavities, making the sensor suitable for mass production.
What problem does this paper attempt to address?