A 4.0-Μvrms 7.5-Khz-bw Neural Signal Recording Circuit for Closed-Loop Deep Brain Stimulation

Ruizhi Tian,Fei Zhou,Qingjiang Xia,You,Wengao Lu,Yacong Zhang,Zhongjian Chen
DOI: https://doi.org/10.1109/icicm59499.2023.10365813
2023-01-01
Abstract:This work presents a low-noise biosignal acquisition application-specific integrated circuit (ASIC) for neural signal recording in closed-loop deep brain stimulation (DBS). An ac-coupled low-noise amplifier (LNA) amplifies the signals of interest while removing the low-frequency electrode offset. An improved programmable-gain amplifier (PGA) employs a flip-over-capacitor structure to prevent the degeneration of frequency response from switches’ off-state resistance. The differential-difference amplifier (DDA) used in the PGA isolates the switching noise from the following switched-capacitor low-pass filter that defines the upper limit of the bandwidth. A 12-bit analog-to-digital converter (ADC) converts the signals to digital representations with a sampling rate of 21.5 kS/s. The proposed circuit was fabricated in a 180 nm CMOS process with an area of 0.576 mm 2 , consuming 60 μA under a 1.8-V supply. It achieves a programmable gain from 47.3 dB to 59.1 dB. The measured input-referred noise is 1.74 μV rms (1 Hz–300 Hz) and 3.73 μV rms (300 Hz–7.5 kHz). The ADC used for digitization achieves an ENOB of 10.63 bits.
What problem does this paper attempt to address?