§Ⅰ : トレーニング

M. Serino,K. Kawakita,Yoshihiko Koseki,Yasushi Yamauchi,Tao Yao,Tatsuhiko Arafune,Aki Kamijo,Onishi Tatsuki,Ayaka Matsui,Ryoichi Nakamura,Kazuya Kawamura,Nobuyoshi Otori
DOI: https://doi.org/10.5759/jscas.22.273
2020-01-01
Abstract::Pericardiocentesis is a procedure to puncture the heart under the cardiac tamponade with a puncture needle from outside the body into the pericardial space, collect and drain pericardial fluid, and release the pressure on the myocardium.However, since the puncture site cannot be visually inspected and the heart beats, it is a very difficult procedure that requires precise control of the needle without over-puncturing the myocardium.As a previous study, we have developed a pulsatile heart simulator.However, there was a problem that the simulated heart part was exposed and different from the actual puncture procedure.Therefore, in this study, in order to reproduce the actual puncture procedure, a phantom for the puncture procedure that reproduces the skin and subcutaneous fat from the abdomen to the chest of the patient was manufactured, and a new sensor-embedded type that can measure the puncture force from outside the body.The purpose is to develop a puncture needle device.A puncture procedure was performed on a beating heart simulator to measure the puncture force.We report the penetration timing of the simulated pericardium and the puncture force during penetration from the measured puncture force and performed a quantitative evaluation as a training device.