A Mechanics-Based Model For Simulating The Needle Deflection In Transverse Isotropic Tissue For A Percutaneous Puncture

WANYU LIU,ZHIYONG YANG,SHAN JIANG
DOI: https://doi.org/10.1142/S021951941950060X
2019-11-25
Journal of Mechanics in Medicine and Biology
Abstract:During the percutaneous puncture for robot-assisted brachytherapy, a medical needle is usually inserted into fiber-structured soft tissue which has transverse isotropic elasticity, such as muscle and skin, to deliver radioactive seeds that kill cancer cells. To place the radioactive seeds more accurately, it is necessary to assess the effect of the transverse isotropic elasticity on the needle deflection. A mechanics-based model for simulating the needle deflection in transverse isotropic tissue is developed in this paper. The anisotropic needle–tissue interaction forces are estimated and used as inputs to drive the model for simulating needle deflections for different insertion orientation angles. Automatic insertion experiments were performed on a single-layered porcine muscle at five different insertion orientation angles. The results show that the maximum difference in the tip deflection for the different insertion orientation angles is 2.99mm when the insertion depth is 50mm. The maximum simulated error of the needle axis deflection is 0.62mm for all insertion orientation angles. The developed model can successfully simulate the needle deflections inside transverse isotropic tissue for different insertion orientation angles. This work is useful for predicting and compensating for the deflection error for automatic needle insertion.
engineering, biomedical,biophysics
What problem does this paper attempt to address?