Estimation Models for Robot-Assisted Needle Insertion

Dedong Gao,Haojun Zheng,Yiping Gao
DOI: https://doi.org/10.1166/asl.2011.1342
2011-01-01
Advanced Science Letters
Abstract:Robot-assisted needle insertion can improve the accuracy of needle placement in minimal invasive surgery. Soft tissue deformation, needle deflection and relative sliding of multi-layer soft tissue directly impact the affects of needle insertion precision. An improved cantilever beam model is proposed to estimate needle deflection. Furthermore, the position of any point on the needle can be calculated in real time with quasi-static thinking. A quasi-static finite element (FE) approach is presented to predict the amount of tissue deformation. With superposed element method, the two and three dimensional quasi-static finite element models are created by ANSYS software and the soft tissue deformation is simulated. From the results of simulation, the models are proved valid and feasible for predicting the location of the needle and the amount of soft tissue deformation. Based on the estimation models, an algorithm is proposed to plan and update the needle path for robot-assisted needle insertion.
What problem does this paper attempt to address?