Breast Intervention Device for Low-Field MRI with a Customized Unilateral Coil

Qingyun Liu,Kecheng Yuan,Qing Zhang,Xueyan Song,Yufu Zhou,Bensheng Qiu
DOI: https://doi.org/10.1016/j.jmr.2023.107579
IF: 2.734
2023-01-01
Journal of Magnetic Resonance
Abstract:With the incidence of breast cancer rising to the top among female malignant tumors, magnetic resonance images guided breast biopsy intervention and minimally invasive treatment have developed as a clinically practical research issue. High field studies have shown the diagnostic value of breast MRI, but the examination costs greatly exceed those of competing conventional mammography. In this case, low-field MRI cannot merely provide typical MRI contrast, but also significantly reduce the cost of diagnosis and treatment for breast cancer patients. This work describes a unilateral breast coil and prototype intervention device, which provides a customized solution for low-field MRI-guided breast intervention. Results demonstrate that the low-field MRI breast intervention device facilitates medical intervention procedures. And the designed positioning device can locate the target lesion within 2-3 mm accuracy. Phantom tests with the customized unilateral coil indicate that the open loops perform as well as the 4-channel commercial closed breast coil, presenting a relatively good SNR (signal-to-noise ratio) and uniformity characteristics. MR scanning images of the volunteer breast using the breast intervention coil also show high SNR, which lays a foundation for further implementation of image-guided breast interventional minimally invasive surgery with the low-field MRI system.
What problem does this paper attempt to address?