Research on Biosensor for Low Concentration Glucose Determination

Zhi‐Xin Guo
2009-01-01
Abstract:A biosensor for the determination of low concentration glucose was prepared by immobilizing enzyme on the thin-film Au electrode surface modified redox polymer.The analytical characteristics were compared by different immobilized enzyme methods including photo polymerization and glutaraldehyde cross-linking.The biosensor fabricated by glutaraldehyde cross-linking showed higher sensitivity and better stability.Glucose biosensor exhibited analytic sensitivity of 35.4 nA μM-1 cm-2 in the relevant concentration range of 0~500 μM at the response time 30 s.The limit of detection for glucose was 0.5 μM.The coefficient of variation was 4.4%.
What problem does this paper attempt to address?