Abstract:<p>Implant-associated infection (IAI) induced by <em>methicillin-resistant Staphylococcus aureus</em> (MRSA) is a devastating complication in the orthopedic clinic. Traditional implant materials, such as Ti6Al4V, are vulnerable to microbial infection. In this study, we fabricated a copper (Cu)-containing titanium alloy (Ti6Al4V-Cu) for the prevention and treatment of MRSA-induced IAI. The material characteristics, antibacterial activity, and biocompatibility of Ti6Al4V-Cu were systematically investigated and compared with those of Ti6Al4V. Ti6Al4V-Cu provided stable and continuous Cu<sup>2+</sup> release, at a rate of 0.106 mg/cm<sup>2</sup>/d. Its antibacterial performance against MRSA <em>in vitro</em> was confirmed by plate counting analysis, crystal violet staining, and scanning electron microscopic observations. Reverse transcription quantitative polymerase chain reaction (RT-qPCR) analysis demonstrated that Ti6Al4V-Cu suppressed biofilm formation, virulence, and antibiotic-resistance of MRSA. The <em>in vivo</em> anti-MRSA effect was investigated in a rat IAI model. Implants were contaminated with MRSA solution, implanted into the femur, and left for 6 weeks. Severe IAI developed in the Ti6Al4V group, with increased radiological score (9.6 ± 1.3) and high histological score (10.1 ± 1.9). However, no sign of infection was found in the Ti6Al4V-Cu group, as indicated by decreased radiological score (1.3 ± 0.4) and low histological score (2.3 ± 0.5). In addition, Ti6Al4V-Cu had favorable biocompatibility both <em>in vitro</em> and <em>in vivo</em>. In summary, Ti6Al4V-Cu is a promising implant material to protect against MRSA-induced IAI.</p>
engineering, biomedical,materials science, biomaterials