Contents determination of four metallic elements in lactate hemofiltration displacement solution by flame atomic absorption spectrometry
JingXian LI,ChunXia JIANG,ShuJuan PIAO,WenQuan LU,WanSheng CHEN
DOI: https://doi.org/10.5428/pcar20170515
2017-01-01
Pharmaceutical Care and Research
Abstract:Objective:To establish a flame atomic absorption spectrometry (FAAS) for contents determination of four me-tallic elements ,namely potassium ,calcium ,magnesium and sodium in the lactate hemofiltration displacement solution (LHDS) . Methods :Potassium ,calcium ,magnesium and sodium ions were determined at the frequencies of 766 .5 ,422 .7 ,285 .2 and 589 .0 nm by FAAS ,using strontium chloride as the masking agent for sodium and magnesiu ,and lanthanum oxide as the mask-ing agent for calcium .Results:Potassium was in good linearity within the ranges of 0 .2029-1 .6228 μg/ml(r=0 .9995) ,with an average recovery rate of (98 .1 ± 0 .98)% .Calcium was in good linearity within the ranges of 5 .0704-15 .2100 μg/ml(r=0 .9999) ,with an average recovery rate of (100 .0 ± 1 .80)% .Magnesium was in good linearity within the ranges of 0 .0800-0 .4000 μg/ml(r=0 .9996) ,with an average recovery rate of (100 .4 ± 1.60)% .Sodium was also in good linearity within the ranges of 0 .2044-1 .6356 μg/ml(r=0 .9993) ,with an average recovery rate of (100 .7 ± 1 .91)% .Conclusion:The method was simple ,rapid ,accurate ,stable and reproducible ,Which could be used for quality control of LHDS .