Determination of anions in the water reducing agent by ion chromatograph

Binhe ZHU,Chengzhu NI,Xunyan ZHAO,Weiqiang GUO,Yan ZHU
DOI: https://doi.org/10.3785/j.issn.1008-9497.2016.03.012
2016-01-01
Abstract:A novel method is presented for the simultaneous determination of gluconate ,chloride ,sulfate and thio‐sulfate in the water reducing agent using ion chromatography with suppressed conductivity detector was eastab‐lished .The sample is diluted with water and filtered through C18 pretreatment column before injection .The separa‐tion is performed on shodex 52‐4e anion exchange column with 3 .6 mmol · L-1 Na2 CO3 at flow rate of 0 .8 mL · min-1 .The sample injection volume is 7 μL .Under the optimum conditions ,it shows the good linear relationships (r≥0 .999 1) between the mass concentration and the peak areas when the concentrations of gluconate ,chloride ,sulfate and thiosulfate were in the range of 5 .0 to 100 .0 ,0 .1 to 30 .0 ,1 .0 to 50 .0 ,1 .0 to 50 .0 mg · L-1 ,respectively .The re‐covery rate ranged from 80 .6% to 113 .9% with the relative standard deviations (RSDs) within 2 .5% (n=7) .What's more ,detection limits calculated at S/N=3 are within 50 .23 and 13 .44μg · L-1 .The method has been applied to detect the anions in the water reducing agent ,which shows high selectivity ,sensitivity and accuracy ,meanwhile easy to implement .
What problem does this paper attempt to address?