Determination of standard value of nitrate standard solution by ion cromatography

Zhang Liping,Xing Zhi,Feng Lu,Li Heng
DOI: https://doi.org/10.3969/j.issn.1672-7916.2006.05.011
2006-01-01
Abstract:The method for exact determination of nitrate in standard solution by ion chromatography with conductivity detector has been developed.The separation of the nitrate was conducted on IonPac AS19 anion exchange column(2mm I.D.x 250mm) with NaOH solution as isocratic eluent,the detection was performed by the suppressed conductivity detector with the injection volume of 25 μL.The optimal chromatography conditions were obtained by orthogonal design L_9(3~4) table.The reliability of method was proved by evaluation of the stability of calibration curve,the test of quality control sample and the recovery.The linearity range of calibration curve is 1.0~100mg/L.The detection limit(3σ),RSD and average recovery of standard addtion are 0.007mg/L,0.36% and 94.2%~104% respectively.
What problem does this paper attempt to address?