Home Monitoring for Clinically Suspected Obstructive Sleep Apnea in Pregnancy.

J. Wang,G. Liu,F. Han
DOI: https://doi.org/10.5664/jcsm.10726
IF: 4.842
2024-01-01
Sleep Medicine
Abstract:Study Objectives: To determine if a home sleep apnea test (HSAT) using a type III portable monitor (PM), Nox-T3 (Nox Medical, Inc., Reykjavik, Iceland), detects obstructive sleep apnea in pregnant women. Methods: Ninety-two pregnant women (34.5 +/- 4.3 years; gestational age 25.4 +/- 8.9 weeks; body mass index 29.9 +/- 4.7 kg/m(2)) with suspected obstructive sleep apnea underwent HSAT with the Nox-T3 PM followed by overnight polysomnography (PSG) and PM recording simultaneously in the laboratory within 1 week. PMs were scored automatically and manually using a 3% criteria and compared with PSGs scored by following guidelines. Results: Apnea-hypopnea indexes were 8.56 +/- 10.42, 8.19 +/- 13.79, and 8.71 +/- 14.19 events/h on HSAT, in-laboratory PM recording, and PSG (P =.955), respectively. Bland-Altman analysis of the apnea-hypopnea index on PSG vs HSAT showed a mean difference (95% confidence interval) of 20.15 (21.83, 1.53); limits of agreement (+/- 2 SD) were 216.26 to 16.56 events/h. Based on a threshold apnea-hypopnea index >= 5 events/h, HSAT had 91% sensitivity, 85% specificity, 84% positive-predictive value, and 92% negative-predictive value compared with PSG. When comparing the simultaneous recordings, closer agreement was observed. Automated vs manual analysis of PM showed no significant difference. Conclusions: A type III PM had an acceptable failure rate and high diagnostic performance operating as a reasonable alternative for in-laboratory PSG in pregnant women.