Validation of the type Ⅲ portable monitor for diagnosis of obstructive sleep apnea in patients with chronic obstructive pulmonary disease

张伟,陈燕文,高莹卉,董霄松,左玉花,余永波,冯军军,常远,韩芳
DOI: https://doi.org/10.3760/cma.j.cn112137-20200226-00471
2020-01-01
Abstract:Objective:To evaluate the utility of a type Ⅲ portable monitor to diagnose obstructive sleep apnea (OSA) in adults with chronic obstructive pulmonary disease (COPD).Methods:A total of 75 COPD patients with suspected OSA, who visited the Peking University International Hospital from September 2015 to November 2018, were recruited and performed home sleep apnea testing (HSAT) using type Ⅲ portable monitor at home followed by an polysomnography (PSG) and the same portable monitor test (PMT) in sleep laboratory. The success rates of HSAT by portable monitor were calculated; the agreement of apnea and hypopnea index (AHI) between HSAT and PSG as well as the PMT and PSG was analyzed by Bland-Altman plot, the sensitivity and specificity of portable monitor in the diagnosis of OSA in COPD patients were evaluated.Results:The success rates of HSAT was 88% (66/75). AHI [ M ( Q1, Q3)] was 16.25 (9.23, 30.68) and 17.95(8.63, 40.50) events/h on HSAT and PMT, respectively, both had no significant difference with AHI on PSG [18.25 (7.48, 44.65) events/h]. Bland-Altman analysis indicated that AHI on HSAT and PMT had high agreement with AHI derived from PSG ( r=0.898, 0.966, both P<0.001). Using an AHI≥30 events/h to diagnose OSA, type Ⅲ portable monitor both at home and in sleep lab had diagnostic specificity over 95%. Conclusion:Type Ⅲ portable monitor can be used either at home or in sleep lab to diagnose OSA in COPD patients.