Transition edge sensor based detector: from X-ray to $γ$-ray
Shuo Zhang,Jing-Kai Xia,Tao Sun,Wen-Tao Wu,Bing-Jun Wu,Yong-Liang Wang,Robin Cantor,Ke Han,Xiao-Peng Zhou,Hao-Ran Liu,Fu-You Fan,Si-Ming Guo,Jun-Cheng Liang,De-Hong Li,Yan-Ru Song,Xu-Dong Ju,Qiang Fu,Zhi Liu
DOI: https://doi.org/10.48550/arXiv.2204.00010
2022-04-27
Abstract:The Transition Edge Sensor is extremely sensitive to the change of temperature, combined with the high-Z metal of a certain thickness, it can realize the high energy resolution measurement of particles such as X-rays. X-rays with energies below 10 keV have very weak penetrating ability, so only a few microns thick of gold or bismuth can obtain quantum efficiency higher than 70\%. Therefore, the entire structure of the TES X-ray detector in this energy range can be realized in the microfabrication process. However, for X-rays or gamma rays from 10 keV to 200 keV, sub-millimeter absorber layers are required, which cannot be realized by microfabrication process. This paper first briefly introduces a set of TES X-ray detectors and their auxiliary systems built by ShanghaiTech University, then focus on the introduction of the TES $\gamma$-ray detector, with absorber based on an sub-millimeter lead-tin alloy sphere. The detector has a quantum efficiency above 70\% near 100 keV, and an energy resolution of about 161.5eV@59.5keV.
Instrumentation and Detectors,Instrumentation and Methods for Astrophysics