An Automated Impedance Enhancement Biopotential Recording Front-End Based on Mode Change Calibration Loop

Lulu Wang,Xinguo Wang,Huayue Song,Yongwei Lou,Yan Liu
DOI: https://doi.org/10.1109/primeasia56064.2022.10103999
2022-01-01
Abstract:Dry electrode interface provides convenient means for continuous biopotential recording, however it requires front-ends with high input impedance to maintain high SNR. This paper proposes a mode change calibration loop (MCCL) for automated input impedance enhancement of capacitively-coupled instrumentation amplifiers (CCIA). To compensate all parasitic capacitors seen at the input nodes which limit the input impedance performance, the MCCL precisely adjusts positive feedback capacitors and automatically converges towards the ideal value. Active shielding is also integrated for cancelling parasitic capacitors outside the chip. Implemented in a typical $\boldsymbol{0.18}-\mu \mathbf{m}$ CMOS process, the post simulation result shows that the front-end consumes $\boldsymbol{4}\ \mu\mathbf{A}$ and achieves $\boldsymbol{1.8}\ \mu\mathbf{V}$ rms input-referred noise over 0.5-100 Hz. After calibration, $\boldsymbol{20}\ \mathbf{G}\Omega$ input impedance at 50 Hz was achieved.
What problem does this paper attempt to address?