Preparation of Fe−Co−LDH/MXene Modified Electrode for Sensitive Determination of Arsenic (III) in Aquatic System

Luyi Chen,Ping Xiao,Bin Hu,Yiwen Wang,Jianmao Yang,Jianyun Liu
DOI: https://doi.org/10.1002/elan.202200521
IF: 3.077
2023-01-01
Electroanalysis
Abstract:3-dimensional (3D) Fe-Co-LDH/MXene composite was synthesized by in-situ synthesis and assembly of Fe-Co-LDH rod around MXene under hydrothermal condition. Due to the unique 3D configuration and good conductivity, the obtained Fe-Co-LDH/MXene modified glassy carbon electrode (Fe-Co-LDH/MXene/GCE) showed excellent electrochemical activity for As(III) detection. Via square-wave anodic stripping voltammetry, the response current on Fe-Co-LDH/MXene/GCE had good linear relationship with As(III) concentrations (1 similar to 1000 ppt) with superior sensitivity (0.22 mu A ppt(-1) cm(-2)) and low detection limit (0.9 ppt). The mechanism of As(III) adsorption was demonstrated. The electrode showed excellent anti-interference ability. Real water sample analysis demonstrated the Fe-Co-LDH/MXene/GCE was deployable in aqua-system.
What problem does this paper attempt to address?