Cause Analysis of Cardiac Tamponade Induced by Intracardiac Echocardiography-Guided Ventricular Arrhythmia Ablation

李赛男,梁明,孙鸣宇,金志清,丁建,张萍,王祖禄,韩雅玲
DOI: https://doi.org/10.3760/cma.j.cn113859-20220920-00203
2022-01-01
Abstract:Objective:To investigate the incidence and causes of cardiac tamponade in intracardiac echocardiography (ICE) guided ventricular arrhythmia (VA) ablation.Methods:Thirty patients underwent ICE guided VA ablation in General Hospital of Northern Theater Command from January 2017 to July 2021 were enrolled.The clinical characteristics, incidence and severity of cardiac tamponade were recorded.The possible causes of cardiac tamponade were analyzed.Results:Among 32 procedures performed in 30 patients [18 male and 12 female, 52.00(23.75, 59.75) years old], cardiac tamponade occurred in 3 patients (9.4%, 3/32). There were 200, 250 and 100 ml volume of blood aspirated separately and no surgical interventions were needed.Ventricular extrasystoles caused by mechanical stimulation of the ICE catheter in the right ventricle were recorded during mapping or ablation in these 3 cases.The 3 mm tip of ICE catheter was invisible under fluoroscopy.Conclusion:The incidence of cardiac tamponade in ICE guided VA ablation was relatively high.The lack of visualization of ICE catheter tip on chest X-ray, the slightly delayed visualization of ICE catheter image on CartoSound system, and weak right ventricular free wall might be the possible causes of cardiac tamponade.
What problem does this paper attempt to address?