Picosecond Time-Resolved X-ray Ferromagnetic Resonance Measurements at Shanghai Synchrotron Radiation Facility

Yang Xia,Cao Jie-Feng,Li Jun-Qin,Zhu Fang-Yuan,Yu Rui,He Jian,Zhao Zi-Long,Wang Yong,Tai Ren-Zhong
DOI: https://doi.org/10.1007/s41365-022-01037-7
2022-01-01
Nuclear Science and Techniques
Abstract:An experimental picosecond time-resolved X-ray ferromagnetic resonance(TR-XFMR) apparatus with a time resolution of 13 ps(RMS) or 31 ps(FWHM)was constructed and demonstrated in the 07U and 08U1A soft X-ray beamlines at the Shanghai Synchrotron Radiation Facility(SSRF) using pump-probe detection and X-ray magnetic circular dichroism(XMCD) spectroscopy. Element and time-resolved ferromagnetic resonance was excited by continuous microwave phase-locking of the bunch clock within the photon beam during synchrotron radiation and was characterized by detecting the magnetic circular dichroism signals of the elements of interest in the magnetic films. Using this equipment, we measured the amplitude of the element-specific moment precession during ferromagnetic resonance(FMR) at 2 GHz in a single Ni81Fe19layer.
What problem does this paper attempt to address?