Photon-in/photon-out Endstation for Studies of Energy Materials at Beamline 02B02 of Shanghai Synchrotron Radiation Facility

Guoxi Ren,Nian Zhang,Xuefei Feng,Hui Zhang,Pengfei Yu,Shun Zheng,Deng Zhou,Zongwang Tian,Xiaosong Liu
DOI: https://doi.org/10.1088/1674-1056/ab5d04
2020-01-01
Chinese Physics B
Abstract:A new photon-in/photon-out endstation at beamline 02B02 of the Shanghai Synchrotron Radiation Facility for studying the electronic structure of energy materials has been constructed and fully opened to users. The endstation has the capability to perform soft x-ray absorption spectroscopy in total electron yield and total fluorescence yield modes simultaneously. The photon energy ranges from 40 eV to 2000 eV covering the K-edge of most low Z-elements and the L-edge of 3d transition-metals. The new self-designed channeltron detector allows us to achieve good fluorescence signals at the low photon flux. In addition, we synchronously collect the signals of a standard reference sample and a gold mesh on the upstream to calibrate the photon energy and monitor the beam fluctuation, respectively. In order to cross the pressure gap, in situ gas and liquid cells for soft x-ray absorption spectroscopy are developed to study the samples under realistic working conditions.