Comparison Between Microporous and Nanoporous Orthodontic Miniscrews : an Experimental Study in Rabbits.

Yueh-Tse Lee,Eric Jein-Wein Liou,Sinn-Wen Chen
DOI: https://doi.org/10.1007/s00056-022-00398-3
2024-01-01
Abstract:Purpose Surface characteristics of orthodontic miniscrews might affect survival rates and removal torque values (RTVs). This experimental study aimed to clarify whether and why a microporous or nanoporous surface promotes higher survival rates and RTVs for orthodontic miniscrews. Methods Using a split-leg design, one set each of nonporous (sham control, n = 24) and microporous (control, n = 6), and three sets of nanoporous (experimental, n = 6 per set) miniscrews were implanted in the tibias of 12 New Zealand rabbits and immediately loaded with 1.5 N nickel-titanium coil springs for 12 weeks. The surface morphology, micropores, and nanotube diameters of the miniscrews were examined using scanning electron microscopy and field-emission scanning electron microscopy. The surface composition and thickness were determined using Auger electron spectroscopy. The survival rates and RTVs of each set were assessed. Results The nanoporous miniscrews had higher survival rates, RTVs (p < 0.001), and thicker nanotube oxide thicknesses (p < 0.001) than the nonporous and microporous miniscrews. The nonporous and microporous miniscrews had no nanotube structures. The surface oxide composition was titanium dioxide (TiO2). The threshold RTV, TiO2 thickness, and nanotube diameter of nanoporous miniscrews needed to promote the experimental survival rate to 100% was determined to be 6.6 +/- 0.8 N-cm (p < 0.05), 22.5 +/- 4.8 nm (p < 0.05), and 17.6 +/- 2.3 nm or above, respectively. Conclusion Nanoporous surfaces promoted higher survival rates and RTVs than microporous miniscrews. This could be due to TiO2 nanotube structures with thicker oxide layers in nanoporous miniscrews.
What problem does this paper attempt to address?