Spectral Calibration Method for Stokes Polarimeter

Guodong Zhou,Jianhui Li,Yanqiu Li,Yuanhe Li
DOI: https://doi.org/10.1117/12.2603827
2021-01-01
Abstract:An efficient method that allows spectral calibration for a Stokes polarimeter is presented in this paper. The calibration procedure only uses a reference polarizer to generate an arbitrary linear polarization state. The reference polarizer could also be calibrated while calibrating the Stokes polarimeter. In addition, this method does not involve the small-angle approximation and could avoid the influence of the initial azimuth error of the retarder and the reference polarizer. The experiment was carried out by a spectroscopy and a polarization state analyzer with a rotating wave plate. Experimental results show that the alignments of axis as well as the true retardation at a specific wavelength of the retarder are obtained.
What problem does this paper attempt to address?