High-energy High-Dose Microfocus X-ray Computed Tomography Driven by High-Average-current Photo-Injector.
Dongcai Hu,Zheng Zhou,Jianxin Wang,Dexin Xiao,Kui Zhou,Peng Li,Shigen Li,Lijun Shan,Hanbin Wang,Yu Liu,Xuming Shen,Chenglong Lao,Xing Luo,Tianhui He,Peng Zhang,Longgang Yan,Jie Liu,Yushou Ding,Zhe Cai,Lei Li,Chengxin Zhang,Qinghua Liu,Jing Li,Yuan Wang,Xingfan Yang,Ming Li,Dai Wu,Menxue Chen,Jianheng Zhao
DOI: https://doi.org/10.3233/xst-210960
2022-01-01
Journal of X-Ray Science and Technology
Abstract:High-energy, high-dose, microfocus X-ray computed tomography (HHM CT) is one of the most effective methods for high-resolution X-ray radiography inspection of high-density samples with fine structures. Minimizing the effective focal spot size of the X-ray source can significantly improve the spatial resolution and the quality of the sample images, which is critical and important for the performance of HHM CT. The objective of this study is to present a 9 MeV HHM CT prototype based on a high-average-current photo-injector in which X-rays with about 70μm focal spot size are produced via using tightly focused electron beams with 65/66μm beam size to hit an optimized tungsten target. In digital radiography (DR) experiment using this HHM CT, clear imaging of a standard 0.1 mm lead DR resolution phantom reveals a resolution of 6 lp/mm (line pairs per mm), while a 5 lp/mm resolution is obtained in CT mode using another resolution phantom made of 10 mm ferrum. Moreover, comparing with the common CT systems, a better turbine blade prototype image was obtained with this HHM CT system, which also indicates the promising application potentials of HHM CT in non-destructive inspection or testing for high-density fine-structure samples.