Study of hybrid micropattern gaseous detector with CsI photocathode for Super Tau-Charm facility RICH

Ping Li,Shiping Chen,Qian Liu,Zhiyong Zhang,Anqi Wang,Jianxin Feng,Baolin Hou,Hongbang Liu,Lei Zhao,Ming Shao,Jianbei Liu,Yangheng Zheng
DOI: https://doi.org/10.1088/1748-0221/18/04/P04028
2023-01-01
Journal of Instrumentation
Abstract:Super tau-Charm facility (STCF) is a future electron-positron collider operating in the tau-Charm energy region with the aim of studying hadron structure and spectroscopy. The baseline design of the STCF barrel particle identification (PID) detector, which covers momentum up to 2 GeV/c, is provided by a Ring Imaging Cherenkov Counter (RICH). The RICH features an approximately focusing design with liquid perfluorohexane sealed in a quartz container as the radiator and a hybrid combination of CsI-coated THGEMs and Micromegas as the photo-electron detector. A 16 x 16 cm2 prototype with a quartz radiator has been built and tested at DESY and stably operated with an effective gain of 105. In this paper, the design, performance, and reconstruction algorithm of RICH detectors are discussed.
What problem does this paper attempt to address?