Hardware Implementation and Evaluation of Electromagnetic Flow Tomography System for Water-Continuous Flows.

Kai Gao,Ziqiang Cui,Zihan Xia,Huaxiang Wang
DOI: https://doi.org/10.1109/tim.2021.3134722
IF: 5.6
2022-01-01
IEEE Transactions on Instrumentation and Measurement
Abstract:Electromagnetic flow tomography (EMFT) technique is a recently developed tomographic method, aiming to map the velocity profiles of water-continuous flows. By measuring the induced electromotive force (emf) on the boundary electrodes, the velocity profile of the conducting fluids moving in the imposed magnetic field could be estimated. Currently, it is still assumed that the conductivity distributions of moving fluids have an insignificant influence on the induced emf. Therefore, the uniform model is the most commonly utilized model in describing the EMFT forward problem. An EMFT measurement system has been designed and implemented, which could accommodate 4 excitation coils and 16 measuring electrodes. In this article, the nonuniform model has been verified by both numerical simulations and experiments, indicating that the presence of high-conductivity inhomogeneity has significant influences on the boundary potentials.
What problem does this paper attempt to address?