Calcium Quantitation with A Parallel EELS/Cooled CCD System

Ruoya Ho,Zhifeng Shao,Andrew P. Somlyo
DOI: https://doi.org/10.1017/s0424820100163940
1996-01-01
Abstract:The value of electron energy loss spectroscopy (EELS) for analyzing low concentrations of phosphorus and calcium, in biological cryo-sections, has been well documented. With currently available techniques, the sensitivities of EELS for phosphorus and calcium can be three to ten fold higher than those of electron probe microanalysis (EPMA).Our experiments were performed with an EELS system including a Philips CM200 electron microscope equipped with a field emission gun and a STEM unit, and a GATAN 666 PEELS. We replaced the original fiber optically coupled photodiode array on the PEELS with a cooled CCD camera with lens coupling for better performance. The microscope is operated at 200 keV, allowing us to use relatively thick cryo-sections (λi = 1700 Å for carbon at 200 keV). We also used the shortest camera length in the STEM unit for achieving higher collection angle. According to our experience, the 2 mm aperture on the PEELS provides the best compromise of energy resolution and collection angle. Under these conditions, the collection semi-angle of this system is about 7 mrad and the energy resolution (full width at half maximum without specimen) is about 2.2~2.4 eV. The main limitations on using a larger collection angle are the 90 mm column between the viewing screen of the microscope and the entrance aperture of the PEELS and the poor energy resolution obtained with larger PEELS aperture.